Penn Medicine moved high-risk nonstress tests into patients’ homes

Utilization of a wireless monitoring device to perform nonstress tests in high-risk pregnancies from home

In short

Women with high-risk pregnancies are asked to come in for nonstress tests (NSTs) again and again near the end of pregnancy. Penn Medicine tested whether they could do those clinically ordered tests at home instead, with the INVU belt and an app that lets the clinician read each test and reply in real time.

Why it was studied

Repeated NSTs in the clinic add a heavy burden to high-risk pregnancies, and the authors note that they widen gaps in care. INVU had already been validated for the baby’s and the mother’s heart rate and for contractions. This study asked whether patients could use it for their real, clinically ordered tests from home.

How it was done

A prospective study at one hospital, Penn Medicine in Philadelphia, from December 2020 to March 2022. 34 women from 32 weeks, carrying one baby, who needed regular fetal testing and had Wi-Fi at home. Reasons for testing included age (14 women), a BMI of 40 or more (11), gestational diabetes (10), chronic high blood pressure (3) and antiphospholipid syndrome (2). Their average BMI was 35.8.

Women put the belt on themselves at home, in place of the clinic test. Clinicians read each NST in a secure app and could message the patient in real time. Patients also rated the system on the System Usability Scale.

What they found

  • Getting started: 5 of the 34 did not complete device training. The other 29 had 131 home NST appointments.
  • Readable tests: 123 appointments (93.9%) gave a test clinicians could interpret; 121 of those (98.3%) were reactive, the reassuring result.
  • No trip: patients avoided a clinic visit in 88.5% of appointments. Only 2 appointments (1.5%) ended in a recommendation for non-urgent delivery, both for raised blood pressure and neither because of the NST.
  • Safety: one patient had mild, passing soreness where a sensor sat, with no redness or bruising.
  • Patients: 23 completed the survey, with a mean score of 76.5 of 100 (“good” usability); 22 of them (95.7%) would prefer home NSTs in a future pregnancy.

Why it matters

The earlier papers showed that INVU measures heart rate and contractions as well as hospital equipment. This one shows patients using it for their actual tests, at home, with nearly nine in ten appointments needing no trip. What patients said about testing at home follows the Penn patients’ views in more detail.

What it does not show

One hospital, 34 women and no comparison group: the study describes what happened and was not designed to show that home tests are as safe as clinic tests or change outcomes. It is not a randomised trial. Women needed Wi-Fi and had to complete training, which 5 did not.

Who ran it

Led by maternal-fetal medicine researchers at Penn Medicine, with Rebecca Hamm as first author. Funded by Nuvo and by a University of Pennsylvania research core (the Community Engagement and Research Core of the Institute for Translational Medicine and Therapeutics). Nuvo took part in design, data collection, analysis, interpretation and writing; three authors work for Nuvo, and Dr. Schwartz is a medical advisor and consultant to Nuvo. Approved by Penn’s Institutional Review Board.

Cover photo: Tako Tsiklauri, Pexels (cropped and extended).

93.9%of home appointments gave a test clinicians could read
88.5%of appointments finished without a clinic visit
95.7%of patients surveyed would choose home tests next time
The paper
Utilization of a wireless monitoring device to perform nonstress tests in high-risk pregnancies from home

Hamm RF, Shkolnik K, Keren N, Reches A, Purnell J, McCabe M, Parry S, Schwartz N. Am J Obstet Gynecol 2023;229(4):463–464.

doi.org/10.1016/j.ajog.2023.06.012
Read in the journal → Copied
Summary by Nuvo. The full text is published by American Journal of Obstetrics & Gynecology.

Highlighted research

Three interwoven rings: People, Process and Place, with safety, trust and experience at the centre Programme studyFront Digit Health · 2026

Hospitalised at home: how Sheba redesigned care for high-risk pregnancy

Sheba’s hybrid unit lets women with high-risk pregnancies be hospitalised at home and visit twice a week. A 2026 study of 40 interviews looks at what changed; INVU provides the fetal heart tracing.

Read the highlight →
Sheba Medical Center’s curved glass building under a blue sky Pilot studyIJGO · 2025

Every other week from home: Sheba’s hybrid care for gestational diabetes

At Sheba, 20 women with gestational diabetes alternated weekly clinic and home visits, with INVU for the nonstress test. 38 of 39 home visits were complete, taking 59 minutes against 159 in clinic.

Read the highlight →
Speech bubbles of different sizes, colours and shapes, each with a different kind of voice, on Nuvo dark green Research letterAJOG MFM · 2024

In their own words: what Penn patients said about nonstress tests at home

Penn asked high-risk patients who did their nonstress tests at home with INVU how it went. They rated it acceptable, valued skipping clinic trips, and their comments on belt fit changed the device.

Read the highlight →
A desert highway at dusk Letter to the editorsAJOG · 2024

A year of remote NSTs in rural care: Banner Health’s view

After a year of INVU in its rural telemedicine programme, a Banner Health obstetrician wrote to AJOG: most remote NSTs were readable, and patients stopped missing visits.

Read the highlight →
Twelve monitor screens, six lit with different real INVU signals ReviewBMC Pregnancy Childbirth · 2023

Fetal monitoring 4.0: an independent review picks out INVU as the device built for home

Researchers in Czechia reviewed fetal monitoring for the post-COVID era. Of six certified monitors they compared, they describe INVU as the only one built for home monitoring rather than labour.

Read the highlight →
How INVU reads contractions: the mother’s heart signal, picked up as biopotential and acoustic signals, swells with each contraction Validation studyAJOG · 2022

Improved uterine contraction detection compared with conventional TOCO

In 118 women in labour, INVU detected 84.8% of the contractions confirmed by the intrauterine catheter, beat the external TOCO belt (82% vs 38%) and kept working in women with obesity.

Read the highlight →
The contraction trace rises and falls, and the height of each of the mother’s heartbeats swells with every contraction Validation studyFront Bioeng Biotechnol · 2022

Reading contractions in the mother’s own heartbeat

The engineering paper behind INVU’s contraction reading: as the uterus tightens, the mother’s heartbeat signals change shape. In 41 women in labour, the method found 89.8% of catheter-confirmed contractions.

Read the highlight →
The mother's heartbeat and the baby's in one INVU recording: tall slow beats and small fast beats Validation studyAJOG MFM · 2020

INVU heart-rate readings agreed closely with hospital CTG

147 women at four centres wore INVU and the hospital monitor at the same time. Fetal heart rate agreed to within about 8 to 9 beats per minute, with no ultrasound and no repositioning.

Read the highlight →